Horiba Ocma 350

Posted on  by 



Horiba Quartz Cell Cap for the OCMA-350. The Quartz Cell is designed to allow full light transmission and to be very clear. Quartz is unique in that it doesn’t absorb UV light like some glasses and plastics would. It also doesn’t react with the majority of samples, especially water and oil samples. Horiba Quartz Cell for the OCMA-350. The Quartz Cell is designed to allow full light transmission and to be very clear. Quartz is unique in that it doesn’t absorb UV light like some glasses and plastics would. It also doesn’t react with the majority of samples, especially water and oil samples.

Measurement principle:

Solvent extraction, NDIR
analysis (Infrared spectroscopy)

Measurement range:

Water:0 to 200 mg/L, Soil:0 to 1000 mg/kg, Parts:0 to 1 Abs.

Resolution

mg/L;

0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg/L

mg/kg;

0 to 9.99 mg/kg; 0.01 mg/kg
10.0 to 99.9 mg/kg; 0.1 mg/kg
100 to 1000 mg/kg; 1 mg/kg

Abs;

0 to 1000 Abs; 0.001 Abs

Repeatability

mg/L;

0 to 9.9 mg/L; ±0.4 mg/L ±1 digit
10.0 to 99.9 mg/L; ±2.0 mg/L ±1 digit
100 to 200 mg/L; ±4 mg/L ±1 digit

Abs;

±1% of full scale

Measurement

Manually controlled

Calibration

Automatic calibration after the calibration standard is introduced to the instrument

Extraction Solvent

S-316 solvent

Sample/Solvent Volume

Approx. 6.5 mL, ratio of sample to solvent is 1:1

Display

Measured value; 3 1/2digits LCD with back-light
Message; Character display LCD with back-light (16 x 2 characters)

Functions

Self diagnostics, Auto hold function, Calendar clock

Output

RS-232C and centronics printer port

Ambient Temperature

0 to 40 ℃

Power

100 to 240 V AC ±10%,50/60 Hz, 60 VA

Dimensions

200 (H) x 250 (W) x 285 (D) mm

Mass

Approx. 5 kg

Ocma

The OCMA-310 uses HORIBA’s special S-316 Solvent to extract the oil components from oily water samples, soil samples, or product surfaces. The extract is mea-sured using IR absorbance, a non-dispersive infrared spectrophotometric technique which is specific to hydrocarbons such as oil. The OCMA-300 provides swift, highly accurate analysis of the oil content of both fresh water and saltwater as well as that of such fluids as those found at factories and wastewater treatment plants. Compact and simple, HORIBA's OCMA-350 oil content analyzer delivers quick, accurate measurements in the lab or in the field.

Horiba Ocma-350 Manual

Compact and simple, HORIBA’s OCMA-350 oil content analyzer delivers quick, accurate measurements in the lab or in the field. It efficiently measures the level of oil in water or soil for environmental applications, checks for residual oil on semiconductor parts which have been cleaned, and measures oil on any industrial surface.